Series Resistance Compensation

Series Resistance Compensation

Series resistance is the sum of all resistances between the amplifier and the inside of the cell using the whole-cell recording method. Due to Ohms Law, the larger this resistance, the greater the difference between the command level and the measured values. This creates an error in actual voltage or current measurement potentially leading to inaccurate observations. To overcome this, the Molecular Devices amplifiers have built-in circuitry to improve the bandwidth of the recording by compensating the error introduced by the voltage or current drop across the series resistance.